Parameters GCMS-QP2020
发布时间:2017-07-01 11:26:18


微信图片_20180727102134

The importance of high-performance analytical instruments for monitoring microscopic quantities of compounds related to environmental pollution and human health, and for developing and evaluating new, highly functional materials and chemical products continues to grow. The GCMS-QP2020 has been designed to meet these needs.
AG视讯 Featuring enhanced instrument functionality, analysis software, databases, and a sample introduction system, the GCMS-QP2020 will help maximize the capabilities of your laboratory.

The vacuum exhaust system, which features a new differential exhaust turbomolecular pump, achieves the highest sensitivity under all GCMS conditions. In addition to helium, it can be operated with confidence using hydrogen or nitrogen as the carrier gas, which helps reduce operational costs. In addition, analysis times can be shortened while retaining high speed and high sensitivity, which contributes to the optimization of laboratory productivity.

AG视讯 The GCMS Insight software package dramatically improves the sensitivity and efficiency of multicomponent batch analysis. The Smart SIM method creation program creates methods to improve the sensitivity in multicomponent analysis. Further, LabSolutions Insight dramatically improves the efficiency of the data processing.

AG视讯 In addition to conventionally used mass spectra, qualitative analysis is supported by three types of valuable information: retention indices, molecular weight information from CI, and accurate mass*. Even higher level qualitative analysis can be performed by combining this information with mass spectra.

With a sample injection system that matches the sample form and the concentrations of the components being measured, as well as a 2D chromatography system that makes full use of the high-speed scan capability, you can configure optimal analysis systems to suit a variety of your analytical needs. And since all these analysis systems are supported by Shimadzu you can put your mind at ease when using them.

AG视讯 The instrument is a high-performance gas chromatograph mass spectrometer with advanced flow control system. At least three injection ports and four detector units can be installed on the same instrument at the same time, and the mass of the detector is not included. Equipped with a microcomputer workstation for chromatograph control and chromatographic data processing.

Partial technical parameters of gas chromatography      

 Gas chromatograph host           

AG视讯      Column Temperature Box: Room Temperature + 4-450     

     Heating rate: at least 220 C /min  /   20 gradient /21 platform program heating       

AG视讯   temperature stability: at least 0.01 degrees centigrade

The cooling rate : 450 to 50 degrees <230 seconds (22 degrees centigrade).        

GCMS-QP2020 has the function of automatic protection of the temperature of the column box           

Keyboard setting program heating           

Supports dual column dual flow dual injection system to improve analysis efficiency.   

AG视讯  Has the ability to upgrade to multi-dimensional gas phase, using two different polar chromatographic columns to separate samples.

Has the technology of upgraded detector shunting system, which can divert the material separated from the analysis column to different detectors.

AG视讯 Two sets diverting / non distributary capillary column inlet    

AG视讯 Flow range: 0~200 mL/min nitrogen, 0~1250 mL/min hydrogen or helium.

AG视讯 The highest temperature of the inlet: 400 C      

AG视讯 Pressure setting range: 0 ~ 149psi          

AG视讯 Pressure setting accuracy: the whole pressure setting ranges from 0 to 149.999 psi in at least 0.001psi.         

   Liquid automatic sampler: in line with dual channel dual injection system requirements          

   Retention time reproducibility: <0.008% or <0.0008min;       

Parameters of mass spectrometer         

 Basic performance of mass spectrometry host  

Range of mass: 2~1090 u

Sensitivity:

EI Scan (helium): 1Pg   

eight fluorine naphthalene OFN m/z 272 S/N > 2000 (RMS)          

EI Scan (hydrogen): 1Pg   eight fluorine naphthalene OFN m/z 272 S/N > 300 (RMS)         

IDL (SIM): IDL < 10 FG (100 FG OFN 8 continuous injection, 272m/z, peak area RSD 3.4%)

Resolution: R > 2M (FWHM)         

  Quality stability: less than 0.1u/48 hours (Heng Wen)          

  Maximum scanning speed: 20000 u/sec      

  Sampling frequency: up to 100Hz       

AG视讯   Iron source material: overall inert high sensitivity ion source designed for patented shielding board        

  Quality analyzer :

A high precision quadrupole with a pre quadrupole.

AG视讯 Equipped with pre quadrupole high precision all metal quadrupole, pre quadrupole can be rotated and cleaned and polished. Main quadrupole can be cleaned and polished, effectively anti pollution.    

 With ASSP rapid scan patent technology, it can minimize the sensitivity of fast scanning and reduce the correctness of mass spectrometry.  

 Scanning function:           

AG视讯 Support full scan mode (Scan), selective ion scanning mode (SIM) and Scan/SIM simultaneous scanning mode.          

  In SIM mode, maximum support for 64 channel x 128 groups.            detecting system       

  The two electron multiplier is equipped with a patent deflection lens (Overdrive Lens) and a + 10kV converter.     

  Off-axis continuous manipulation of electronic multiplier            Dynamic range: 8 x 106

Vacuum system      

AG视讯     High vacuum: dual entry differential turbine molecular pump exhaust system, 200L/sec + 200L/sec, the total pumping rate is greater than 400L/sec.    

    Low vacuum: 30L/min (60Hz) mechanical pump        

    Standard is equipped with a series of vacuum gauge and ion gauge (software for direct monitoring of high vacuum and low vacuum).         

   The maximum column flow can reach 15mL/min (He), which can directly connect the column with the maximum 0.53mm inner diameter.         

   It supports double column dual flow system (Twin Line system), and the two column flow control systems adopt advanced flow control unit.      

   It supports the use of hydrogen as carrier gas without replacing any components.       

   This mass spectrometry software can control the existing mass spectrum hosts in our unit, and a set of software can control two sets of hosts. The software needs to implement the deconvolution set function.